Product Release Information visionCATS 2.1

Release Number: 2.1.15272.1 (2.1)
Release Date: 29/09/2015

This version has been tested and released on Windows 7 Professional 64-bit and Windows 10 64-bit (see “Known Problems and Workarounds” for detailed information).

Customer Support: For more information or support, please email us at support@camag.com

Introduction

This document contains the release notes for VisionCATS version 2.1. The following sections describe the release in detail and provide late-breaking or other information that supplements the main documentation.

Installation and Upgrade Notes

Installation

See the installation instructions for full details.

IMPORTANT: You must completely uninstall any previous “developer release” or “early access” version of this product before installing this release.

Manifest

This release consists of the following items:

Release notes
Drivers
Instrument Firmware
visionCATS installer
Example Files (you may import them after installation)

Minimum System Requirements

System Memory: 4000MB
Free Disk Space: 100GB
Operating System: Windows 7
Networking: Internet access

What’s new in visionCATS 2.1?

  • Spectrum scanning by using Scanner 4 or Scanner 3 instruments
  • Ability to control the sequence type (Sample or Reference) for each evaluation
  • Scanner advanced mode: when using a scanner step in multi-wavelength mode, it’s now possible to set the lamp, filter and mode for each wavelength
  • Many GUI-related improvements in the quantitative evaluation part
  • Folder backups and restore in the Explorer view
  • Dual wavelength
  • CV value (coefficient of variation) is displayed instead of σ (sigma, Standard deviation) for calibration function and sample results
  • Corrected restore to include scans, wincats analysis and method collection entries

Known Problems and Workarounds

Defect: Exporting a report as Word document does not work.
Workaround: Use another export format.

Defect: Profiles viewer stops responding when two or more viewers is visible at the same time (docking or multi windows)
Workaround: Close the corresponding analysis of the blocked viewer, and reopen it without another profiles viewer visible.

Defect: The import of winCATS methods won’t import steps of instruments not supported in visionCATS, and some winCATS specific parameters will be ignored
Reason: winCATS and visionCATS don’t exactly support the same set of instruments, and for each instrument, the exact same set of parameters
Workaround: After having imported some winCATS methods, visionCATS will inform the user when some steps are ignored. In all cases, please review the imported methods.

Defect: After changing of Visualizer, a Visualizer cannot be executed in an analysis having already images
Reason: Only one Visualizer can be used in an analysis
Workaround: Redo the analysis

Defect: After changing of camera for a Visualizer, already opened analysis with Visualizer step waiting for execution, or Visualizer Manual control or OQ may behave strangely (using old camera, bad image corrections…)
Workaround: Before changing of camera, close all analysis with next step as Visualizer and Visualizer views

Defect: After changing of camera for a Visualizer, Visualizer in toolbox isn’t marked as having a bad Diagnostic
Workaround: Restart the client

Defect: While using two visualizers on a single PC, one or both camera aren’t detected, or the camera of one visualizer seems to be connected to the other.
Workaround: We strongly suggest to use only one Visualizer per InstrumentService

Defect: No image appears in Data or in Comparison after installing for the first time visionCATS on the machine
Workaround: Restart Windows

Defect: Sequence table don’t allow to drag and drop a multi-selection
Workaround: Before clicking for drag and during the drag, hold SHIFT key for dragging a multi-selection

Defect: Applications crashes after clicking on an image or opening a comparison
Workaround: Update your graphic driver to the latest version

Defect: After installation, no visionCATS is installed. Also, the SQLServer service may not want to start.
Workaround: Change the name of the user which install the application, username should be different from machine and domain name

Defect: Instrument aren’t responding correctly
Workaround: Check if you have an Intel(R) Active Management Technology – SOL COM port on your machine, and disable the corresponding COM port in visionCATS/General Settings

Defect: visionCATS display “Out of memory” with only around 1.2GB of 4GB memory used by visionCATS, on a 32bits system
Workaround: increase the applications memory space with ‘bcdedit /set increaseuserva 3072’ command, and restart Windows

Defect: when using offline help, the display of some pages is very slow
Reason: the page requests an online resource that is not accessible, causing the slow down
Workaround: use the online help when possible. If not possible, try using Chrome or Firefox (instead of Internet Explorer), where the slow down occurs less often.

Defect: visionCATS fails in Scanner/Scanner3 Diagnostics if the Positioning-Test is started  in the same session.
Reason: Scanner data of the new test cannot be stored for later processing. This does not invalidate the test results of the previous run.
Workaround: Close the current Scanner Instrument Tab and reopen it. Attention: just switching between the Tabs inside the Instrument Tab will not prevent from the failure. You need to close the Instrument Tab.

Important for Windows 10: For proper working of the Visualizer under Windows 10, the Windows power settings have to be changed as follows: Go to Power Options/Choose what the power button does/Change settings that are currently available/Shutdown settings and disable the option “Turn on fast startup”.

Swiss Marketing Leadership Award 2015 – and the Winner is…CAMAG!

Muttenz, Switzerland – 29 September 2015  – CAMAG was elected out of more than 500 Swiss companies as the winner of the Swiss Marketing Leadership Award 2015. On the occasion of the 15th Swiss CRM Forum, held in Zurich on 15 September 2015, a high profile jury with members from industry and academia chose CAMAG as the winner of the Swiss Marketing Leadership Award 2015, based on a study made by ZHAW School of Management and Law.

The Swiss CRM Forum – the Marketing Leadership Conference – is the leading Swiss community event for innovative customer relationship management and new marketing. The annual one-day event provides a perfect meeting and knowledge platform all leaders with customer focus. The conference conveys marketing trends and perspectives on topics such as marketing leadership, customer relationship management (CRM), customer experience management (CEM) and digital transformation. Lectures, panel discussions and interactive formats, such as best practice cases, workshops and round table discussions provide the participants access to a wide range of information and in-depth subject knowledge.

Translated transcript of the laudatio by Dr. Brian P. Rüeger (Jury President):

“CAMAG was founded in 1958 in Muttenz and has been active in Thin-Layer Chromatography since 1961. CAMAG develops and manufactures both basic and high-end equipment and associated software for all steps of Thin-Layer Chromatography. CAMAG prevailed against many top performers in the Swiss marketing leadership study. The study will be released today and is available for free. By showing best practice marketing leadership on the performance as well as on the customer level, CAMAG has managed to achieve world leadership in a difficult market. On the performance level they have constantly expanded their offering with innovative products and evolutionarily maintained the existing product range. So they are revolutionary and evolutionary at the same time. On the customer level, they have pushed forward the customer acquisition and (think of the strong Swiss franc) successfully maintained existing customers under difficult conditions through geographic expansion and the development of new industries. At the same time, you are not only revolutionary and evolutionary, but also explorers and survival artist. CAMAG has prevailed, because the company exemplary works the four marketing areas which create value for the company. This is nothing less than marketing leadership. Congratulations.”

CAMAGflash August 2015

The HPTLC Method Library

USP has just published the Dietary Supplement Compendium (DSC) 2015. For HPTLC users worldwide this is a milestone because two general chapters ‹203› and ‹1064› are included. These chapters define and explain parameters to run standardized HPTLC. Both chapters become official in August of this year.

CAMAG’s new generation software, visionCATS, provides a Method Library with methods that are in full compliance with ‹203›. All methods that have been added to this edition of DSC are in the library. Also included are methods of identification from the European Pharmacopoeia and methods published by the International Association for the Advancement of HPTLC. Other methods will be added continually.

To read the complete CAMAGflash, please download the PDF below.

Product Release Information visionCATS 2.0 SP2

Release Number: 2.0.15169.1 (2.0 SP2)
Release Date: 06/07/2015
Customer Support: For more information or support, please visit our website or email us at support@camag.co

Introduction

This document contains the release notes for visionCATS version 2.0 SP2. The following sections describe the release in detail and provide late-breaking or other information that supplements the main documentation.

Installation and Upgrade Notes

Installation

See the installation instructions for full details.

IMPORTANT: You must completely uninstall any previous “developer release” or “early access” version of this product before installing this release.

Manifest

This release consists of the following items:

Release notes (this file)
Drivers
Instrument Firmware
visionCATS installer
Example Files (you may import them after installation)

Minimum System Requirements

System Memory: 4000MB
Free Disk Space: 100GB
Operating System: Windows 7
Networking: Internet access

What’s new in visionCATS 2.0?

  • Integration of the TLC Scanner 3 and TLC Scanner 4 instruments
  • Support of multiple quantitative evaluations with the scanner instrument
  • Support of multiple image based evaluations with the visualizer
  • 21CFR11 compliance
  • Import of winCATS data in a viewing mode
  • Support of multiple application steps
  • Possibility to define report templates
  • Supports electronic signatures
  • SST extended to scanner data
  • Online help

What’s new in visionCATS 2.0 SP2?

  • Corrected backup to include scans, wincats analysis and method collection entries, and corrected rights issue with backup to folder
  • Corrected report generation when evaluation’s smoothing, baseline or peak detection is disabled
  • Included area information in export peak, and removed condition to have the “Data Export” option in order to perform an export
  • Display a correct message when importing a corrupted license key
  • Scanner OQ Monochromator test: Text Bandwith -> Bandwidth; New calculation for 20nm Bandwidth
  • Linomat 5 no longer skips tracks
  • Correction in 1.3-1.4 migration
  • Scanner OQ crash when scanner OQ completed, or when positioning test is started a 2nd time
  • winCATS import: corrected several bugs
  • Data view: Corrected an error when track sequencer goes outside of the image

Known Problems and Workarounds

Defect: winCATS import will remove any DigiStore, Reportstar or spectrum steps, without any error messages
Reason: DigiStore and Reportstar aren’t supported by visionCATS, and version 2.0 also doesn’t have spectrum support
Workaround: For documentation steps, add a Visualizer, for methods with spectrum, import them in visionCATS 2.1

Defect: The display in visionCATS of the version of the scanner firmware is incorrect
Reason: there is a conversion problem between hex and dec coding. 1.14.30 corresponds to 1.20.48

Defect: After changing of Visualizer, a Visualizer cannot be executed in an analysis having already images
Reason: Only one Visualizer can be used in an analysis
Workaround: Redo the analysis

Defect: After changing of camera for a Visualizer, already opened analysis with Visualizer step waiting for execution, or Visualizer Manual control or OQ may behave strangely (using old camera, bad image corrections…)
Workaround: Before changing of camera, close all analysis with next step as Visualizer and Visualizer view

Defect: After changing of camera for a Visualizer, Visualizer in toolbox isn’t marked as having a bad Diagnostic
Workaround: Restart the client

Defect: While using two visualizers on a single PC, one or both camera aren’t detected, or the camera of one visualizer seems to be connected to the other.
Workaround: We strongly suggest using only one Visualizer per Instrument Service

Defect: No image appears in Data or in Comparison after installing for the first time visionCATS on the machine
Workaround: Restart Windows

Defect: Sequence table don’t allow to drag and drop a multi-selection
Workaround: Before clicking for drag and during the drag, hold SHIFT key for dragging a multi-selection

Defect: Applications crashes after clicking on an image or opening a comparison
Workaround: Update your graphic driver to the latest version

Defect: After installation, no visionCATS is installed. Also, the SQLServer service may not want to start.
Workaround: Change the name of the user which install the application, username should be different from machine and domain name

Defect: Instrument aren’t responding correctly
Workaround: Check if you have an Intel(R) Active Management Technology – SOL COM port on your machine, and disable the corresponding COM port in visionCATS/General Settings

Defect: visionCATS display “Out of memory” with only around 1.2GB of 4GB memory used by visionCATS, on a 32bits system
Workaround: increase the applications memory space with ‘bcdedit /set increaseuserva 3072’ command, and restart Windows

Defect: An error occur while importing a visionCATS 2.0 analysis with evaluation
Reason: The evaluation improvement of SP1 has changed some data type, migration is only done on system update, not on import
Workaround: export again the analysis from a visionCATS 2.0 SP1 system

Defect: visionCATS fails in Scanner/Scanner3 Diagnostics if the Positioning-Test is started  in the same session. 
Reason: Scanner data of the new test cannot be stored for later processing. This does not invalidate the test results of the previous run.
Workaround: Close the current Scanner Instrument Tab and reopen it. Attention: just switching between the Tabs inside the Instrument Tab will not prevent from the failure. You need to close the Instrument Tab.

CAMAGflash June 2015

New case study: Determination of monoacylglycerides in biodiesel

Last year CAMAG has started to develop case studies focusing on typical analytical tasks which can be performed in a faster, easier, cheaper and/or better way by HPTLC. In addition, application videos are provided for practical support. The first case studies dealt with typical applications from the herbal medicines and food application field. The latest two case studies deal with industrial applications. The here mentioned case study “Determination of monoacylglycerides in biodiesel” describes the use of a fast AMD gradient to separate a minor compound from the fatty-acid methyl esters in biodiesel. Please also have a look at the case study about “Cleaning validation at API production units” performed within less than one hour. We are continuously providing our customers with new HPTLC applications to be downloaded at our website. Using visionCATS allows you to access our constantly expanding HPTLC Method Library.

To read the complete CAMAGflash, please download the PDF below.

CAMAGflash April 2015

Case Study: Quantitative determination of steviol glycosides

The adage “A picture is worth a thousand words” ideally refers to the image-giving planar chromatographic method for the evaluation of samples. In this case study, an HPTLC method for analysis of steviol glycosides with its single steps was visualized in a short video recorded at the JLU Giessen, Germany. Only sugar-containing analytes, and so the steviol glycosides, were detected in separated raw extracts, Stevia formulations, or sugar-free food products using a selective derivatization with the ß-naphthol reagent. Coupling to mass spectrometry was employed for confirmation of the identity. The workflow of the method can be conveyed more effective as a visualized experiment than as a pure description. Enjoy the video and agree with the updated adage “A video is worth a thousand words”.

To read the complete CAMAGflash, please download the PDF below.

Product Release Information visionCATS 2.0 SP1

Release Number: 2.0.15069.1 (2.0 SP1)
Release Date: 10/03/2014
Customer Support: For more information or support, please email us at support@camag.com

Introduction

This document contains the release notes for VisionCATS version 2.0 SP1. The following sections describe the release in detail and provide late-breaking or other information that supplements the main documentation.

Installation and Upgrade Notes

Installation

See the installation instructions for full details.

IMPORTANT: You must completely uninstall any previous “developer release” or “early access” version of this product before installing this release.

Manifest

This release consists of the following items:

Release notes
Drivers
Instrument Firmware
visionCATS installer
Example Files (you may import them after installation)

Minimum System Requirements

System Memory: 4000MB
Free Disk Space: 100GB
Operating System: Windows 7
Networking: Internet access

What’s new in visionCATS 2.0?

  • Integration of the TLC Scanner 3 and TLC Scanner 4 instruments
  • Support of multiple quantitative evaluations with the scanner instrument
  • Support of multiple image based evaluations with the visualizer
  • 21CFR11 compliance
  • Import of winCATS data in a viewing mode
  • Support of multiple application steps
  • Possibility to define report templates
  • Supports electronic signatures
  • SST extended to scanner data
  • Online help

What’s new in visionCATS 2.0 SP1?

  • Evaluation’s concentration can now be entered with unit of choice, fg/ml is now possible
  • Corrected several minors bugs in evaluation
  • DNS queries for client/server communications have now fallback in case of unknown host
  • Slow network will now not remove edited lock
  • Corrected instrument progress display
  • Improved some sections of help    

Known Problems and Workarounds

Defect: The display in visionCATS of the version of the scanner firmware is incorrect.
Reason: There is a conversion problem between hex and dec coding. 1.14.30 corresponds to 1.20.48.

Defect: After changing of Visualizer, a Visualizer cannot be executed in an analysis having already images.
Reason: Only one Visualizer can be used in an analysis.
Workaround: Redo the analysis.

Defect:After changing of camera for a Visualizer, already opened analysis with Visualizer step waiting for execution, or Visualizer Manual control or OQ may behave strangely (using old camera, bad image corrections…).
Workaround: Before changing of camera, close all analysis with next step as Visualizer and Visualizer views.

Defect: After changing of camera for a Visualizer, Visualizer in toolbox isn’t marked as having a bad Diagnostic.
Workaround: Restart the client.

Defect: While using two visualizers on a single PC, one or both camera aren’t detected, or the camera of one visualizer seems to be connected to the other.
Workaround: We strongly suggest to use only one Visualizer per Instrument Service.

Defect: No image appears in Data or in Comparison after installing for the first time visionCATS on the machine.
Workaround: Restart Windows.

Defect: Sequence table doesn’t allow to drag and drop a multi-selection.
Workaround: Before clicking for drag and during the drag, hold SHIFT key for dragging a multi-selection.

Defect: Applications crashes after clicking on an image or opening a comparison.
Workaround: Update your graphic driver to the latest version.

Defect: After installation, no visionCATS is installed. Also, the SQLServer service may not want to start.
Workaround: Change the name of the user which installs the application, username should be different from machine and domain name.

Defect: Instruments are not responding correctly.
Workaround: Check if you have an Intel(R) Active Management Technology – SOL COM port on your machine, and disable the corresponding COM port in visionCATS/General Settings.

Defect: visionCATS display Out of memory with only around 1.2GB of 4GB memory used by visionCATS, on a 32bit system.
Workaround: increase the applications’ memory space with “bcdedit /set increaseuserva 3072 command, and restart Windows.

Defect: An error occur while importing a visionCATS 2.0 analysis with evaluation
Reason: The evaluation improvement of SP1 has changed some data type, migration is only done on system update, not on import
Workaround: export again the analysis from a visionCATS 2.0 SP1 system.

CAMAGflash January 2015

visionCATS 2.0 – quantitative evaluation of HPTLC chromatograms

While developing visionCATS, we have incorporated more customer feedback than in any previous CAMAG TLC software. Through extensive tests, the designs were fine-tuned to enable intuitive handling and maximize user friendliness. In visionCATS 2.0 we have included new features that make analysts’ lab work easier while meeting the needs of cGMP compliant routine work and keeping the flexibility for method development in scientific research. From visionCATS 2.0 you can expect the very best software for qualitative and quantitative HPTLC analysis.

To read the complete CAMAGflash, please download the PDF below.

Product Release Information visionCATS 2.0.0

Release Number: 2.0.0
Release Date: 19/12/2014

Customer Support: For more information or support, please email us at support@camag.com

Introduction

This document contains the release notes for visionCATS version 2.0.0. The following sections describe the release in detail and provide late-breaking or other information that supplements the main documentation.

Installation and Upgrade Notes

Installation

See the installation instructions for full details.

IMPORTANT: You must completely uninstall any previous “developer release” or “early access” version of this product before installing this release.

Manifest

This release consists of the following items:

Release notes
Drivers
Instrument Firmware
visionCATS installer

Minimum System Requirements

System Memory: 4000MB
Free Disk Space: 100GB
Operating System: Windows 7
Networking: Internet access

What’s new in visionCATS 2.0?

  • Integration of the CAMAG scanner3 and scanner4 instruments
  • Support of multiple quantitative evaluations with the scanner instrument
  • Support of multiple image based evaluations with the visualizer
  • 21CFR11 compliance
  • Import of winCATS data in a viewing mode
  • Support of multiple application steps
  • Possibility to define report templates
  • Supports electronic signatures
  • SST extended to scanner data
  • Online help

Known Problems and Workarounds

Defect: The display in visionCATS of the version of the scanner firmware is incorrect.
Reason: There is a conversion problem between hex and dec coding. 1.14.30 corresponds to 1.20.48.

Defect: After changing of Visualizer, a Visualizer cannot be executed in an analysis having already images.
Reason: Only one Visualizer can be used in an analysis.
Workaround: Redo the analysis.

Defect:After changing of camera for a Visualizer, already opened analysis with Visualizer step waiting for execution, or Visualizer Manual control or OQ may behave strangely (using old camera, bad image corrections…).
Workaround: Before changing of camera, close all analysis with next step as Visualizer and Visualizer views.

Defect: After changing of camera for a Visualizer, Visualizer in toolbox isn’t marked as having a bad Diagnostic.
Workaround: Restart the client.

Defect: While using two visualizers on a single PC, one or both camera aren’t detected, or the camera of one visualizer seems to be connected to the other.
Workaround: We strongly suggest to use only one Visualizer per Instrument Service.

Defect: No image appears in Data or in Comparison after installing for the first time visionCATS on the machine.
Workaround: Restart Windows.

Defect: Sequence table doesn’t allow to drag and drop a multi-selection.
Workaround: Before clicking for drag and during the drag, hold SHIFT key for dragging a multi-selection.

Defect: Applications crashes after clicking on an image or opening a comparison.
Workaround: Update your graphic driver to the latest version.

Defect: After installation, no visionCATS is installed. Also, the SQLServer service may not want to start.
Workaround: Change the name of the user which installs the application, username should be different from machine and domain name.

Defect: Instruments are not responding correctly.
Workaround: Check if you have an Intel(R) Active Management Technology – SOL COM port on your machine, and disable the corresponding COM port in visionCATS/General Settings.

Defect: visionCATS display Out of memory with only around 1.2GB of 4GB memory used by visionCATS, on a 32bit system.
Workaround: increase the applications’ memory space with “bcdedit /set increaseuserva 3072 command, and restart Windows.

CAMAGflash October 2014

Cumulative CAMAG Bibliography Service (CCBS) database is now online

The most comprehensive compilation of literature in the field of Planar Chromatography (TLC/HPTLC) is the Cumulative CAMAG Bibliography Service (CCBS). It covers more than 11’000 abstracts of TLC/HPTLC publications between 1982 and today. Providing reports was established by Dr. Dieter Jänchen in 1965 and the workflow and form evolved over the years. Manually written and transferred into the printed CBS form, it was set-up as a data-base in 1997. With CBS 93 in 2004, the electronic database for download was introduced and with the latest CBS 113, the online search. Test it!